logo
Found 90 products for "

insertion tube

"
Qualidade KARL STORZ 11278A1 ureteroscopo para reparação Fábrica

KARL STORZ 11278A1 ureteroscopo para reparação

KARL STORZ 11278A1 ureteroscope for repair ► Brief Description Fault: 1) Image spotted with partially black point, image bundle aged 2) Light bundle partially wire broken 3) Insert tube aged 4) Biopsy tube abrasion and leakage 5) Wires and coil pipe abrasion Solution: Replace HD image bundle, light bundle, insert tube, wires, coil pipe and rubber. ►Problem & Solutoin Brand KARL STORZ Model 11278A1 Fault 1) Image spotted with partially black point, image bundle aged 2) Light

Qualidade Insighters IS3-F laringoscope para reparação Fábrica

Insighters IS3-F laringoscope para reparação

Insighters IS3-F laryngoscope for repair ► Brief Description Fault: 1. Image blank screen; 2. Insert tube abrasion 3. Bending section transformation; 4. Distal head transformation; Solution: Replace CCD, insert tube, bending section, rubber ►Problem & Solutoin Brand Insighters Model IS3-F Fault 1. Image blank screen; 2. Insert tube abrasion 3. Bending section transformation; 4. Distal head transformation; Solution Replace CCD, insert tube, bending section, rubber ► Common

Qualidade SonoScape EG-500 Gastroscopia Eletrônica para Reparação Fábrica

SonoScape EG-500 Gastroscopia Eletrônica para Reparação

SonoScape EG-500 Electronic Gastroscopy For Repair ► Brief Description Inspection Results: 1. No image display, CMOS water ingress fault; 2. 20% of the light bundle is broken; 3. There are bite marks on the insertion tube in multiple places; 4. The biopsy tube is worn out and not smooth; 5. The angle is seriously insufficient and tight, with wear on the snake bone, steel wire, and helix tube; 6. The light guide tube has dents; 7. The handle shell is cracked and has severe

Qualidade OLYMPUS TJF-Q180V Duodenoscópio Eletrônico Para Reparação Fábrica

OLYMPUS TJF-Q180V Duodenoscópio Eletrônico Para Reparação

OLYMPUS TJF-Q180V Electronic Duodenoscope For Repair ► Brief Description Inspection Results: 1. Image display is available; 2. The light bundle brightness is acceptable; 3. The insertion tube is worn; 4. The light guide tube is worn; 5. The angle is insufficient, the snake bone, steel wire and helix tube are worn; 6. The forceps elevator handle is damaged; 7. The handle shell is severely cracked. Solution: Replacement: Insertion tube, handle shell, light guide tube, snake

Qualidade Toracoscópio Olympus LTF-160 para reparo Fábrica

Toracoscópio Olympus LTF-160 para reparo

Olympus LTF-160 thoracoscope for repair ► Brief Description Fault: 1.with image, but big risk for CCD after leakage. 2. Light bundle broken completely. 3. Insert tube with little scratch. 4. Biopsy tube abrasion and leakage. 5. Wires and helix tube abrasion. 6. Grip broken, leakage. 7. All the buttons failure. 8. Rubber leakage. 9. the whole scope leakage. Solution: Replace CCD,light bundle, biopsy tube, wires,helix tube,grip and rubber, repair all the buttons. ►Problem &

Qualidade R Wolf 7326071 ureteroscópio para reparação Fábrica

R Wolf 7326071 ureteroscópio para reparação

R Wolf 7326071 ureteroscope for repair ► Brief Description Fault: The image has spots,image bundle aging,light bundle broken;Insert tube aging;Biopsy channel worn and leakage;Bending section,wire and helix tube were worn. Solution: Replace the light bundle,image bundle,rubber,biopsy channel,helix tube,wire. ►Problem & Solutoin Brand R Wolf Model 7326071 Fault The image has spots,image bundle aging,light bundle broken;Insert tube aging; Biopsy channel worn and leakage;Bending

Qualidade Olympus CYF V2 Cistoscópio para Reparação Fábrica

Olympus CYF V2 Cistoscópio para Reparação

Olympus CYF V2 Cystoscope For Repair ► Brief Description Fault: Leakage. Solution: Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber. ►Problem & Solutoin Brand Olympus Model CYF V2 Fault Leakage Solution Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber ► Common issues with CYF V2 Cystoscope No Image / Dark Image Poor Image Resolution / Blurry Image Ineffective Suction Difficulty

Qualidade Olympus A70940A Artroscópio autoclavável para reparação Fábrica

Olympus A70940A Artroscópio autoclavável para reparação

Olympus A70940A Autoclavable Arthroscope for repair ► Brief Description Fault: 1. Image blur, black dots and dark. 2. Fiber not light enough. 3. Rod lens broken. 4. Objective lens broken and dirty. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Clean and install the optical path, check the image. 8. Seal with endoscope and test. ►Problem & Solutoin Brand Olympus Model

Qualidade GIF-XQ240 Gastroscópio da Olympus para reparação Fábrica

GIF-XQ240 Gastroscópio da Olympus para reparação

GIF-XQ240 Gastroscope by Olympus for repair ► Brief Description Fault: Leakage. Solution: Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ►Problem & Solutoin Brand Olympus Model GIF-XQ240 Fault Leakage. Solution Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ► Common issues with GIF-XQ240 Gastroscope No Image / Dark Image Poor Image Resolution / Blurry Image Ineffective Suction Difficulty with Angulation (Stiff or Loose) Bubbles in the Field of View