Casos de reparação de dispositivos médicos

Qualidade Colonoscópio Olympus CF-H190L para reparação Fábrica

Colonoscópio Olympus CF-H190L para reparação

Olympus CF-H190L Colonoscope for repair ► Brief Description Fault: 1. Image ok; 2. Light bundle still ok; 3. The front end of the insert tube is broken, and the serpentine; the position of stress boot is leaking; 4. Biopsy tube wear; 5. Left Angle failure, bending section, wire and helix tube wear; ...
Qualidade Samsung Medison P2-4AH Sonda cardíaca para reparação Fábrica

Samsung Medison P2-4AH Sonda cardíaca para reparação

Samsung Medison P2-4AH Cardiac probe for repair ► Brief Description Fault:Lens problem, crystal damage. Solution:Replace probe lens, replace crystal. ►Problem & Solutoin Brand Samsung Model P2-4AH Fault Lens problem, crystal damage Solution Replace probe lens, replace crystal ► Common issues with P2...
Qualidade DUODENOSCÓPIO OLYMPUS TJF-Q180V PARA REPARO Fábrica

DUODENOSCÓPIO OLYMPUS TJF-Q180V PARA REPARO

OLYMPUS TJF-Q180V DUODENOSCOPE FOR REPAIR ► Brief Description Fault: 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5. Bendingsection, wires and helix tube wear; 6. Knob leakage. Solution: Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and ...
Qualidade Olympus CF-240L Colonoscópio para reparação Fábrica

Olympus CF-240L Colonoscópio para reparação

Olympus CF-240L Colonoscope for repair ► Brief Description Fault: Leakage, glass of light bundle blur, rubber leak, angle not enough,image fuzzle,objective lens scratch. Solution: Replace rubber, objective lens, C cover. Adjust the angle. ►Problem & Solutoin Brand Olympus Model CF-240L Fault Leakage...
Qualidade Endoscópio Karl Storz 27005BA para reparação Fábrica

Endoscópio Karl Storz 27005BA para reparação

Karl Storz 27005BA endoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...
Qualidade Sonda de matriz faseada Toshiba PST-25BT para reparo Fábrica

Sonda de matriz faseada Toshiba PST-25BT para reparo

Toshiba PST-25BT Phased Array Probe For Repair ► Brief Description Fault: Housing break, image weak, crystal aging. Solution: Replace crystal and housing. ►Problem & Solutoin Brand Toshiba Model PST-25BT Fault Housing break, image weak, crystal aging Solution Replace crystal and housing ► Common ...
Qualidade Cabeça de câmera STRYKER 1188-210-105 para reparo Fábrica

Cabeça de câmera STRYKER 1188-210-105 para reparo

STRYKER 1188-210-105 camera head for repair ► Brief Description Fault: Image with interrupt. Cable aging. Solution: Replace compatible new cable and connector. ►Problem & Solutoin Brand STRYKER Model 1188-210-105 Fault Image with interrupt. Cable aging Solution Replace compatible new cable and ...
Qualidade Cabeça da câmara STORZ H3-ZA para reparação Fábrica

Cabeça da câmara STORZ H3-ZA para reparação

STORZ Camera Head H3-ZA For Repair ► Brief Description Fault:Image streaks frequently,cable loose contact and aged. Solution: Replace cable kit. ►Problem & Solutoin Brand STORZ Model H3-ZA Fault Image streaks frequently,cable loose contact and aged Solution Replace cable kit ► Common issues with H3...
Qualidade GE RIC6-12-D sonda de endocavidade para reparação Fábrica

GE RIC6-12-D sonda de endocavidade para reparação

GE RIC6-12-D endocavity probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens,Clean cable. ►Problem & Solutoin Brand GE Model RIC6-12-D Fault Lens problem Solution Replace lens,Clean cable ► Common issues with RIC6-12-D Probe Dome Damage and Fluid Leaks Image Dropouts ...
Qualidade Hitachi C41V1 sonda de ultra-som para reparação Fábrica

Hitachi C41V1 sonda de ultra-som para reparação

Hitachi C41V1 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Brand Hitachi Model C41V1 Fault Lens problem Solution Replace lens ► Common issues with C41V1 Probe Image Dropouts or Streaks Cable and Connector Damage Acoustic Lens ...
Qualidade Sonda linear Sonoscape 8-L743 para reparo Fábrica

Sonda linear Sonoscape 8-L743 para reparo

Sonoscape 8-L743 linear probe for repair ► Brief Description Fault: Crystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model 8-L743 Fault Crystal problem Solution Replace the crystal ► Common issues with 8-L743 Probe Acoustic Lens Damage Piezoelectric Crystal Damage ...
Qualidade Sonoscope C362 Sonda abdominal para reparação Fábrica

Sonoscope C362 Sonda abdominal para reparação

Sonoscape C362 Abdominal probe for repair ► Brief Description Fault: Cystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model C362 Fault Cystal problem Solution Replace the crystal ► Common issues with C362 Probe Image Dropouts (Black Lines or Streaks) Damaged Lens ...
Qualidade GE 4C-RS sonda abdominal para reparação Fábrica

GE 4C-RS sonda abdominal para reparação

GE 4C-RS abdominal probe for repair ► Brief Description Fault: Lens broken, cable broken, crystal bad. Solution: Replace probe head, repair cable. ►Problem & Solutoin Bran GE Model 4C-RS Fault Lens broken, cable broken, crystal bad Solution Replace probe head, repair cable ► Common issues with 4C-RS ...
Qualidade Sonda de ultrassom Careutech BK Medical 9048 para reparo Fábrica

Sonda de ultrassom Careutech BK Medical 9048 para reparo

BK Medical 9048 Ultrasound Probe For Repair ► Brief Description Fault: The housing break. Solution: Replace probe housings. ►Problem & Solutoin Brand BK Medical Model 9048 Fault The housing break. Solution Replace probe housings. ► Common issues with 9048 Probes No Image / Poor Image Quality Probe ...
Qualidade Peças de equipamento médico Sonoscape L743 sonda linear para reparação Fábrica

Peças de equipamento médico Sonoscape L743 sonda linear para reparação

Sonoscape L743 linear probe for repair ► Brief Description Fault: Probe head broken. Solution: Replace probe head.(including crystal and upper housing). ►Problem & Solutoin Brand Sonoscape Model L743 Fault Probe head broken Solution Replace probe head.(including crystal and upper housing) ► Common ...
Qualidade BK Medical 9096 sonda de ultra-som robótica para reparação Fábrica

BK Medical 9096 sonda de ultra-som robótica para reparação

BK Medical 9096 Robotic Ultrasound Probe For Repair ► Brief Description Fault: All the outer skin of the cable is seriously damaged. Solution: Replace the outer skin of the cable. ►Problem & Solutoin Brand BK Medical Model 9096 Fault All the outer skin of the cable is seriously damaged Solution ...
Qualidade Sonda Samsung Medison 3D4-9 SN P0N7  Para Reparo Fábrica

Sonda Samsung Medison 3D4-9 SN P0N7 Para Reparo

Samsung Medison 3D4-9 Probe SN P0N7 For Repair ► Brief Description Fault: Scanhead broken. Solution: Replace scanhead ► Problem & Solutoin Brand Samsung Model 3D4-9 Fault Scanhead broken Solution Replace scanhead ► Common i ssues with 3D4-9 Probes Cable damage Lens delamination Element failure (3D...
Qualidade Reparação de aparelhos médicos Casas de GE RIC5-9-D sonda de ultra-som para reparação Fábrica

Reparação de aparelhos médicos Casas de GE RIC5-9-D sonda de ultra-som para reparação

GE RIC5-9-D Probe Fault: Dome loose, straight relief tiny broken, with error on machine. Solution: Replace dome and oiling. Repair cost : $***. Common Issues with the GE RIC5-9-D Ultrasound Probe: Image Artifacts/Dropouts: These can manifest as lines, shadowing, or a complete loss of image in ...
Qualidade Fios angulares e tubos helicoidais CF-HQ190L 20% de luz quebrada Fábrica

Fios angulares e tubos helicoidais CF-HQ190L 20% de luz quebrada

Products: CF-HQ190L For Repair Problem & Solutoin: CF-HQ190L - Light bundle 20% Broken Knob Leaked The angle wire and helix tube Worn Biospy channel Worn Rubber Cracked Light guide tube Indented Solution Replace Biopsy Channel, Wire, Light Guide Tube, Knob, Rubber Applications: When it comes to ...
Qualidade Equipamento Médico Casos De Reparo De Olympus CF-H190L Colonoscópio Para Reparo Fábrica

Equipamento Médico Casos De Reparo De Olympus CF-H190L Colonoscópio Para Reparo

Olympus CF-H190L Colonoscope for repair,SN:261****. Fault:Light bundle broken,biopsy tube worn,water air tube leakage,bending section deformation,wire helix tube worn. Solution:Replace the bending section,wire,helix tube,air tube and rubber.PS:Warranty for replace parts only. Repair time:7 working ...
Qualidade Equipamento Médico Olympus reparado Solução definitiva para Serviços Médicos Fábrica

Equipamento Médico Olympus reparado Solução definitiva para Serviços Médicos

Product Description: Product Summary: The Medical Device Repair Cases product offered is a comprehensive solution for repairing and maintaining Olympus medical devices. With a warranty of 90 days, customers can have peace of mind knowing that their repaired devices are covered for an extended period...
Qualidade OLYMPUS CF-HQ190L Caixas de reparação de dispositivos médicos Caixa de ferramentas de reparação Fábrica

OLYMPUS CF-HQ190L Caixas de reparação de dispositivos médicos Caixa de ferramentas de reparação

OLYMPUS CF-HQ190L Repair Problem: 1, Image display;2, Light bundle broken about 10%;3, Insertion tube is ok;4, Biospy channel worn and leaking;5, The upper angle broken, helix tube and bending section were worn;6, Light guide tube indented and the back cover is leaking;7, The EL-connector is not ...